Please use this identifier to cite or link to this item: http://nopr.niscair.res.in/handle/123456789/915
Title: Acrylic acid grafted PVC membrane based ion selective electrode for calcium and hardness measurement of water
Authors: Nanda, D
Oak, M S
Kumar, M Pravin
Issue Date: Feb-2007
Publisher: CSIR
Series/Report no.: Int. Cl.⁸ G01N27/333
Abstract: A poly(vinyl chloride) based membrane ion selective electrode has been developed in which acrylic acid has been grafted on poly(vinyl chloride) powder. The ion selective electrode system, prepared without an internal solution, contains DBP (plasticizer) and NaTPB (anion excluder), and shows near Nernstian response to Mg²⁺, Cu²⁺ and super-Nernstian response to Ca²⁺ in the concentration range 10⁻¹ – 10-5 mol L⁻¹. EDTA has been used to mask the interference of transition metal ions. The best performance of the electrode was obtained when it was stored in 10⁻² mol/L Ca²⁺ solution when the electrode was not in use. The electrode, with response time ≤ 20 s, has been found useful in the determination of concentration of Ca²⁺ and hardness in river water.
Description: 258-262
URI: http://hdl.handle.net/123456789/915
Appears in Collections:IJC-A Vol.46A(02) [February 2007]

Files in This Item:
File Description SizeFormat 
IJCA 46A(2) (2007) 258-262.pdf190.58 kBAdobe PDFView/Open


Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.