Please use this identifier to cite or link to this item: http://nopr.niscair.res.in/handle/123456789/8470
Title: A novel concept in the design of a psychrometer
Authors: Sen, Sunit Kumar
Issue Date: Dec-2005
Publisher: CSIR
IPC Code: G01N25/62
Abstract: A novel psychrometric method for measurement of humidity of a process is proposed. The design of a psychrometer is a complex one due to simultaneous introduction of two variables at the same time, viz., the meter current is dependent on both the relative humidity and ambient temperature. The circuit design has been so made that the said current would depend only on the relative humidity of air and independent of temperature variations at the place of measurement. Theoretical calculations show that the maximum error in the measurement of the meter current would be limited to within +1.6% at 120°F and -2% at 60°F for 100% relative humidity while the same would be within ±1.75% for the entire temperature range 60°-120°F for 0% relative humidity.
Page(s): 498-504
URI: http://hdl.handle.net/123456789/8470
ISSN: 0975-1017 (Online); 0971-4588 (Print)
Appears in Collections:IJEMS Vol.12(6) [December 2005]

Files in This Item:
File Description SizeFormat 
IJEMS 12(6) 498-504.pdf117.46 kBAdobe PDFView/Open