Please use this identifier to cite or link to this item: http://nopr.niscair.res.in/handle/123456789/8298
Title: Determination of full energy peak efficiency of HpGe detector from 59.5 to 1332.5 keV
Authors: Gültekin, Ali
Kaynak, Gökay
Gürler, Orhan
Keywords: Efficiency
HpGe detector
Monte Carlo simulation
Issue Date: Apr-2006
Publisher: CSIR
Series/Report no.: G01T
Abstract: To detect very low radio activities in the environmental samples, it is sometimes necessary to know the full energy peak efficiency of the detector. This work proposes a full energy peak efficiency simulation which represents adequately the high purity germanium coaxial detector responses in the energy range 59.5-1332.5 keV and at distances of 5, 10 and 25 cm
Description: 281-286
URI: http://hdl.handle.net/123456789/8298
ISSN: 0975-1041 (Online); 0019-5596 (Print)
Appears in Collections:IJPAP Vol.44(04) [April 2006]

Files in This Item:
File Description SizeFormat 
IJPAP 44(4) 281-286.pdf213.41 kBAdobe PDFView/Open


Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.