Please use this identifier to cite or link to this item: http://nopr.niscair.res.in/handle/123456789/7065
Title: Complexometric determination of palladium(II) using ethanethiol as a selective masking agent
Authors: Karthikeyan, J
Parameshwara, P
Shetty, A Nityananda
Shetty, Prakash
Keywords: Complexometry
Masking reagent
EDTA titration
Ethanethiol
Palladium determination
Issue Date: Sep-2006
Publisher: CSIR
Series/Report no.: C22B1/244
Abstract: A complexometric method based on the selective masking property of ethanethiol towards palladium(II) is proposed. In presence of diverse metal ions, palladium(II) is complexed with excess EDTA and the surplus EDTA is backtitrated with standard zinc sulphate solution at <i style="">p</i>H 5-5.5 (acetic acid-sodium acetate buffer) using xylenol orange as indicator. An excess of a 0.1% aqueous solution of ethanethiol is then added to displace EDTA from Pd(II)-EDTA complex. The released EDTA is titrated with the same standard zinc sulphate solution as before. Reproducible and accurate results are obtained in the concentration range of 0.5-14.26 mg of palladium with relative error of ±0.40% and coefficient of variation not exceeding ±0.34%. The effect of diverse ions is studied. The method has been used for the determination of palladium in its complexes, catalysts and synthetic alloy mixtures.
Description: 484-487
URI: http://hdl.handle.net/123456789/7065
ISSN: 0975-0991 (Online); 0971-457X (Print)
Appears in Collections:IJCT Vol.13(5) [September 2006]

Files in This Item:
File Description SizeFormat 
IJCT 13(5) 484-487.pdf27.62 kBAdobe PDFView/Open