Please use this identifier to cite or link to this item: http://nopr.niscair.res.in/handle/123456789/6988
Title: A chloride ion-selective potentiometric sensor based on a polymeric schiff base complex
Authors: Kumar, K Girish
John, K Saji
Indira, C J
Keywords: Ion selective electrodes
Chloride ion
Amino methylated polystyrene
Issue Date: Jan-2006
Publisher: CSIR
Series/Report no.: Int. Cl.<sup>7</sup> C07C251/02
Abstract: This paper describes the fabrication of an ion-selective electrode in which a polymeric schiff base complex of cobalt(II) is used as the ionophore. The main advantage of the electrode is that it is mechanically stable upto 3 months. The electrode shows a linear response in the range of 2.5 <img src='/image/spc_char/cross.gif' border=0> 10<sup>-5</sup>-0.5 <img src='/image/spc_char/cross.gif' border=0> 10<sup>-1</sup> mol dm<sup>-3</sup>. The response time of the electrode is 30 s. The pH range at which the electrode works is 3.8 to 6.8. The electrode was found to be selective towards chloride ion in the presence of ions like Na<sup>+</sup>, Ca<sup>2+</sup>, Mn<sup>2+</sup>, Fe<sup>3+</sup>, Co<sup>2+</sup>, Ni<sup>2+</sup>, Cu<sup>2+</sup>, Zn<sup>2+</sup>, CH<sub>3</sub>COO<sup>-</sup>, NO<sub>3</sub><sup>-</sup>, SO<sub>4</sub><sup>2-</sup>, Br<sup>- </sup>and NO<sub>2</sub><sup>-</sup>.
Description: 13-16
URI: http://hdl.handle.net/123456789/6988
ISSN: 0975-0991 (Online); 0971-457X (Print)
Appears in Collections:IJCT Vol.13(1) [January 2006]

Files in This Item:
File Description SizeFormat 
IJCT 13(1) 13-16.pdf120.9 kBAdobe PDFView/Open