Please use this identifier to cite or link to this item:
http://nopr.niscair.res.in/handle/123456789/54017
Title: | Automation for calibrating a precision current source by Ohm’s law method |
Authors: | Ehtesham, Bushra Ushakiran John, Thomas |
Keywords: | Current source;Automated calibration;Ohm’s law method;Uncertainty;LabVIEW |
Issue Date: | Feb-2020 |
Publisher: | NISCAIR-CSIR, India |
Abstract: | We present the design of a program developed in LabVIEW for automating the calibration of a precision current source by the Ohm’s law method. The program in addition to automating the instrument setup and measurement performs uncertainty computations in accordance with guide to the expression of uncertainty in measurement (GUM) and produces the calibration report and saves a record of the calibration process with all relevant information. The program has been tested for calibrating Keithley 6430 remote source meter in the current range of 1nA to 100 mA. While there are several benefits from the automation, but the most significant one is the achievement of low uncertainties in calibrating the current source. |
Page(s): | 99-105 |
URI: | http://nopr.niscair.res.in/handle/123456789/54017 |
ISSN: | 0975-0959 (Online); 0301-1208 (Print) |
Appears in Collections: | IJPAP Vol.58(02) [February 2020] |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
IJPAP 58(2) 99-105.pdf | 514.96 kB | Adobe PDF | View/Open |
Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.