Please use this identifier to cite or link to this item: http://nopr.niscair.res.in/handle/123456789/5289
Title: Design and implementation of controlled drug infusion system
Keywords: Drug infusion
Drop sensing assembly
Microprocessor
Alarms
I V set
Issue Date: Oct-2005
Publisher: CSIR
Series/Report no.: A61M5/14
Abstract: An inexpensive controlled drug infusion system based around 80C85 microprocessor with RAM, EPROM and peripheral interface devices is developed. In this system, the drop rate selection is done by BCD thumb wheel switches from 1 to 99 drops/min, LCD display is used for displaying the parameters settings, number of drops infused and various alarms messages for patient safety and unique drop sensing assembly is incorporated for accurate detection of falling drops in drip chamber. The prototype has been tested successfully on neonates.
Description: 761-766
URI: http://hdl.handle.net/123456789/5289
ISSN: 0975-1084 (Online); 0022-4456 (Print)
Appears in Collections:JSIR Vol.64(10) [October 2005]

Files in This Item:
File Description SizeFormat 
JSIR 64(10) 761-766.pdf238.21 kBAdobe PDFView/Open


Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.