dc.description.abstractPresent paper describes studies carried out on an industrial dead weight tester (pressure range, up to 330 MPa; calibration; 3 times; period, 6 years), one pressure dial gauge (pressure range, up to 70 MPa; calibration, 3 times; period, 3 years), and one digital pressure calibrator (pressure range, up to 200 MPa; calibration, 4 times; period, 6 years). A small change was observed in the short-term stability but the change was prominent in case of long-term stability in all the instruments. Stability of dead weight tester and pressure dial gauge was well within their claimed manufacturer uncertainties while it was found to be beyond claimed uncertainty in case of digital pressure calibrator.en_US
dc.language.isoen_USen_US
dc.publisherCSIRen_US
dc.relation.ispartofseriesG01L1/00en_US
dc.sourceJSIR Vol.65(09) [September 2006]en_US
dc.subjectCalibrationen_US
dc.subjectHydraulic pressure deviceen_US
dc.subjectPressure measurementen_US
dc.titleStudies on the stabilities of various types of industrial pressure measuring devicesen_US