dc.description | 320-321 | en_US |
dc.description.abstract | The present paper discusses the design and development of a force transducer. The force transducer has been developed for a nominal capacity of 5 kN and strain gauges have been applied at suitable locations over it to form a Wheatstone bridge. The force transducer has been metrologically investigated according to the standard calibration procedures based on standards ISO 376-2011. | en_US |
dc.language.iso | en_US | en_US |
dc.publisher | NISCAIR-CSIR, India | en_US |
dc.rights | ![]() | en_US |
dc.source | JSIR Vol.75(05) [May 2016] | en_US |
dc.subject | Force transducer | en_US |
dc.subject | ISO 376-2011 | en_US |
dc.subject | Uncertainty of measurement | en_US |
dc.title | Design, Development and Metrological Characterization of a Force Transducer | en_US |