Please use this identifier to cite or link to this item: http://nopr.niscair.res.in/handle/123456789/34146
Title: Measurements of dose rate for 10-16 keV Synchrotron X-ray photons at BL-16 beam line of Indus-2
Authors: Sinhamahapatra, Debasis
Haridas, G
Kumar, Pradeep
Ghodke, A D
Tiwari, M K
Hannurkar, P R
Keywords: Synchrotron radiation;Absorbed dose rate;Dosimeter
Issue Date: Apr-2016
Publisher: NISCAIR-CSIR, India
Abstract: Indus-2 is a 2.5 GeV electron synchrotron radiation source located at Raja Ramanna Centre for Advanced Technology (RRCAT), India. There is significant radiation exposure in the direct beam of synchrotron radiation due to its intense flux and low energy at synchrotron radiation beam line. In this paper, measurements of absorbed dose rate for mono chromatic synchrotron X-ray photon of energy ranging 10-16 keV is presented using various types of passive dosimeters, like CaSO4(Dy) TLDs, LiF :Mg,Ti (TLD-100), and Gafchromic film EBT-2 at microfocus X-ray fluorescence spectroscopy beam line (BL-16). The measured results are also compared with theoretically calculated values. It is observed that LiF :Mg,Ti (TLD-100), and Gafchromic film EBT-2 show 3-6 times more absorbed dose rates as compared to calculated values whereas CaSO4(Dy) TLD shows 10-20 times more absorbed dose rates.
Page(s): 259-262
URI: http://hdl.handle.net/123456789/34146
ISSN: 0975-1041 (Online); 0019-5596 (Print)
Appears in Collections:IJPAP Vol.54(04) [April 2016]

Files in This Item:
File Description SizeFormat 
IJPAP 54(4) 259-262.pdf144.2 kBAdobe PDFView/Open


Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.