Please use this identifier to cite or link to this item:
Title: Determination of Allan deviation of Cesium atomic clock for lower averaging time
Authors: Banerjee, P
Chatterjee, Arundhati
Keywords: Cesium atomic clock
Allan deviation
Lower averaging time
Issue Date: Dec-2007
Publisher: CSIR
Abstract: Absolute Allan deviation of the Cesium clock for averaging time (τ) of 5 days or more may be calculated from the corresponding data of circular T published by Bureau International des Poids et Mesures (BIPM). For lower values of τ, the Allan deviation may indirectly be found from the extrapolation of these values through τ¹/² fit as recommended by CCTF-WGMRA guidelines. Absolute Allan deviation may also be directly found out by inter-comparison of minimum three clocks assuming that the noise in all clocks is fully uncorrelated. This paper analyses the values of Allan deviation determined by the direct method keeping in mind the limitation of the measurement system. These values of Allan deviation tally well with those found from the data of circular T.
Description: 945-949
ISSN: 0019-5596
Appears in Collections:IJPAP Vol.45(12) [December 2007]

Files in This Item:
File Description SizeFormat 
IJPAP 45(12) 945-949.pdf110.74 kBAdobe PDFView/Open

Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.