Please use this identifier to cite or link to this item: http://nopr.niscair.res.in/handle/123456789/25682
Full metadata record
DC FieldValueLanguage
dc.contributor.authorGupta, Ramesh Chandra-
dc.contributor.authorSingh, S P-
dc.date.accessioned2014-01-15T06:59:46Z-
dc.date.available2014-01-15T06:59:46Z-
dc.date.issued2005-12-
dc.identifier.issn0975-105X (Online); 0367-8393 (Print)-
dc.identifier.urihttp://hdl.handle.net/123456789/25682-
dc.description438-445en_US
dc.description.abstractThis paper presents theoretical analysis of the specific absorption rate (SAR) distribution in three-layered tissue (fat, muscle and lung layers as in thorax region) in direct contact with a box-horn applicator for hyperthermia treatment of cancerous tumor. The box-horn is assumed to be filled with dielectric (ceramic NPOT-96) to provide a better impedance match to bio-media. The present analysis is based on plane wave spectral technique. The spatial distribution of SAR in fat, muscle and lung layers are computed and reported at 433 and 2450 MHz. Wider SAR-distribution in each of the bio-layers in cross-sectional area and higher penetration depth in lung are observed at 433 MHz in comparison to those at 2450 MHz for dielectric-loaded box-horn. The normalized SAR distributions are also reported for various tissues (fat/muscle/lung layers) at the same depth in order to make a comparative study in these tissues. A modified box-horn, in which the horn is flared in both E- and H-planes is also considered. The effect of substituting pyramidal horn in place of H-plane sectoral horn in the design of box-horn on the SAR distribution is also examined. The results have been validated against available published results in the literature.en_US
dc.language.isoen_USen_US
dc.publisherNISCAIR-CSIR, Indiaen_US
dc.relation.ispartof84.40.Ba, 84.40.Azen_US
dc.rights CC Attribution-Noncommercial-No Derivative Works 2.5 Indiaen_US
dc.sourceIJRSP Vol.34(6) [December 2005]en_US
dc.subjectBio-mediaen_US
dc.subjectCanceren_US
dc.subjectDielectric-loaded box-hornen_US
dc.subjectHyperthermiaen_US
dc.subjectPlane wave spectraen_US
dc.subjectSpecific absorption rate (SAR)en_US
dc.titleTheoretical modeling of a dielectric-loaded box-horn applicator for hyperthermia treatment of tumors in thorax regionen_US
dc.typeArticleen_US
Appears in Collections:IJRSP Vol.34(6) [December 2005]

Files in This Item:
File Description SizeFormat 
IJRSP 34(6) 438-445.pdf1.92 MBAdobe PDFView/Open


Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.