Please use this identifier to cite or link to this item: http://nopr.niscair.res.in/handle/123456789/20554
Title: Instrumentation for a retarding potential analyzer for measurement of electron parameters
Authors: Rajput, S S
Garg, S C
Issue Date: Oct-2002
Publisher: NISCAIR-CSIR, India
Abstract: Various design considerations and their implementation in an instrument, designed for measurement of electron parameters on-board Indian satellites SROSS-C2, along its orbital path are presented. The proposed instruments measures the electron density in the range 5x102 - 5x106 cm-3 electron temperatures from 500 K to 4000 K, and photoelectron flux of energy 3-30 eV of the order 108 cm- s-1 with full -scale instrument sensitivity 0.2%. Experimental results obtained during testing and flight demonstrate the instrument performance.
Page(s): 315-322
URI: http://hdl.handle.net/123456789/20554
ISSN: 0975-1017 (Online); 0971-4588 (Print)
Appears in Collections:IJEMS Vol.09(5) [October 2002]

Files in This Item:
File Description SizeFormat 
IJEMS 9(5) 315-322.pdf1.89 MBAdobe PDFView/Open


Items in NOPR are protected by copyright, with all rights reserved, unless otherwise indicated.