Please use this identifier to cite or link to this item: http://nopr.niscair.res.in/handle/123456789/17960
Title: Mercury(II) ion sensing with PVC-matrix based membrane sensors
Authors: Mahajan, Rakesh Kumar
Kaur, Ravneet
Lobana, Tarlok S
Issue Date: Mar-2006
Publisher: NISCAIR-CSIR, India
IPC Code: Int. Cl.8 G01N27/333
Abstract: Thiosemicarbazone derivatives have been used as ion carriers for the preparation of PVC-matrix based mercury(II)selective membrane sensors. The electrodes give near-Nernstian responses in the linear concentration range of 1.0×10-1-5.0×10-6 M  with detection limits of the order of ~10-6 M  . The stable potentiometric signals are obtained within a short time period of 20-255. The effect of different plasticizers has been studied and dioctylsebacate (DOS) found to give a better response in comparison to other plasticizers. Selectivity coefficient values (log KPotHg,M ) have been evaluated using fixed interference method. Better selectivity for mercury(II) ions is observed over many of the mo novalent (Na+, K+ and NH4+) and divalent ions (Mg2+, Ca2+, Zn 2+ , Pb2+ , Ni2+, Co2+, etc.). The sensors have also been used as indicator electrodes in potentiometric titration of mercury(II) ions with EDT A and its determination in synthetic water samples.
Page(s): 639-642
URI: http://hdl.handle.net/123456789/17960
ISSN: 0975-0975(Online); 0376-4710(Print)
Appears in Collections: IJC-A Vol.45A(03) [March 2006]

Files in This Item:
File Description SizeFormat 
IJCA 45A(3) 639-642.pdf830.04 kBAdobe PDFView/Open