Please use this identifier to cite or link to this item: http://nopr.niscair.res.in/handle/123456789/15306
Title: Purification of tryptic peptides for mass spectrometry using polyvinylidene fluoride Membrane
Authors: Kurien, B T
Matsumoto, H
Scofield, R H
Issue Date: Aug-2001
Publisher: NISCAIR-CSIR, India
Abstract: A simple procedure for the purification of tryptic peptides, prior to mass spectrometric analysis, using polyvinylidene fluoride membrane (PVDF) is described. The sensitivity of mass spectrometric analysis is such that minor impurities in tryptic peptide digests suppress the signal obtained. However, we obtained useful signal, from a sample that did not yield any spectra earlier, by purifying the sample using PVDF membrane. For this. the tryptic peptide digest was first spotted on the membrane which was then air-dried and washed. Further, the membrane was extracted with trifluoroacetic acid (TFA) and acetonitrile and subjected to mass spectrometric analysis. This procedure enabled us to identify a cross-reactive D1 antigen on the neutrophil surface that bound antibodies that targeted 60 kD Ro autoantigen in systemic lupus erythematosus, an autoimmune disorder.
Page(s): 274-276
URI: http://hdl.handle.net/123456789/15306
ISSN: 0975-0959 (Online); 0301-1208 (Print)
Appears in Collections:IJBB Vol.38(4) [August 2001]

Files in This Item:
File Description SizeFormat 
IJBB 38(4) 274-276.pdf723.93 kBAdobe PDFView/Open