Please use this identifier to cite or link to this item: http://nopr.niscair.res.in/handle/123456789/1362
Full metadata record
DC FieldValueLanguage
dc.contributor.authorKaur, Jaspreet-
dc.contributor.authorKumar, Jagdish-
dc.date.accessioned2008-05-26T09:12:39Z-
dc.date.available2008-05-26T09:12:39Z-
dc.date.issued2008-05-
dc.identifier.issn0022-4456-
dc.identifier.urihttp://hdl.handle.net/123456789/1362-
dc.description366-370en_US
dc.description.abstractThis paper describes design and development of a gas flow measurement system, which is particularly helpful for airflow measurement in anaesthesia ventilator. The system measured accurately, required range (0-70 l/m), of airflow for anaesthesia ventilator. Main advantages of developed technique are simplicity of design, dynamic stability and fast response time. Thermistor based hardware design associated with graphical approach has been developed successfully and incorporated in indigenous prototype of anaesthesia ventilator.en_US
dc.language.isoen_USen_US
dc.publisherCSIRen_US
dc.sourceJSIR Vol.67(5) [May 2008]en_US
dc.subjectAirflowen_US
dc.subjectAnaesthesia ventilatoren_US
dc.subjectFlowmeteren_US
dc.subjectGas flow rateen_US
dc.subjectMass flow controlleren_US
dc.subjectThermistoren_US
dc.titleDesign and development of thermistor based gas flow measurement system for anaesthesia ventilatoren_US
dc.typeArticleen_US
Appears in Collections: JSIR Vol.67(05) [May 2008]